Post "Ir 4f photoemission spectrum for the element (reseacrh grade)"

forum: Ir 4f

from: Engr. Dulce María Guzmán Bucio

499 Spicer points

1 Followers

Post

Information

  • Ir 4f photoemission spectrum for the element (reseacrh grade) (Ir 4f forum)

    Fitting of the Ir 4f photoemission spectrum for the element (research grade), data were provided by Vincent Crist (XPS library website, https://xpslibrary.com/). The spectrum was fitted employing three Voigt doublets. The branching ratio was forced to be equal to the theoretically expected value (0.7866), which was taken from the Scofield tables. The fit needs to be optimized, since it fails to reproduce the data in the circled region (~50 eV).
    Read more

    Information about the sample:

    • Shape: Other :element

    • Sample material: Ir

    • Elements: Ir

    • Sample composition: Ir

    Data acquisition parameters:

    • Photon source: Al Ka (monochromatic)

    • Pass energy: 14.25

    • Take-off angle: 90°

    Equipment and analysis software:

    • Software used: Analyzer

    xpsoasis_506_image_100.png

Full name:

Dulce María Guzmán Bucio

Title:

Engineer

organization:

Cinvestav

Country:

México

This user is:

Moderator in elements:

Moderator in forums:

Number of post in database:

7

Number of post:

Total: 41

(4 posts ) (4 posts ) (3 posts ) (3 posts ) (3 posts ) (3 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (2 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts ) (1 posts )